The DRSL-U isolated universal
input DIN rail signal conditioner
provides a competitive choice in
terms of both price and technology
for galvanic isolation of process
signals to SCADA systems or
PLC equipment.
The DRSL-U
accepts thermocouple, RTD, linear
resistance, potentiometer, voltage
or current input and converts these
signals to linear voltage or current
output. The unit offers isolation
between input, output and supply,
provides surge suppression and
protects control systems from
transients and noise.
The DRSL-U
also eliminates ground loops and
can be used for measuring floating
signals. Low power consumption
facilitates DIN rail mounting without
the need for any air gap. The
DRSL-U is easily configured by using
the DRSL-DISPLAY programming
interface in conjunction with the
DRSL-ADAPTOR configuration
adaptor.
The DRSL-DISPLAY has
a 4 line LCD display with scrolling
help text in 7 languages (English,
French, German, Italian, Spanish,
Danish and Swedish) which guides
the user through all the configuration
steps. The DRSL-U is designed
with electronic hardware switches,
consequently it is not necessary to
open the device to set any internal
DIP-switches.
SPECIFICATIONS INPUT RTD Input Input Type: Pt10, Pt20, Pt50, Pt100, Pt200, Pt250, Pt300, Pt400, Pt500, Pt1000, Ni50, Ni100, Ni120, Ni1000, linear resistance, potentiometer Cable Resistance Per Wire: 50 Ω max Sensor Current: 0.2 mA nominal Effect of Sensor Cable Resistance (3-Wire/4-Wire RTD): <0.002 Ω / Ω Sensor Error Detection: Yes Short Circuit Detection: <15 Ω Linear Resistance Input Range: 0 to 10,000 Ω Potentiometer Input Range: 10 to 100 kΩ Thermocouple Input Input Types: J, K, T, E, R, S, B, N, L, U, W3, W5, LR Cold Junction Compensation (CJC) Via Internally Mounted Sensor: ±(2.0°C + 0.4°C * Δt) (Δt = internal temperature-ambient temperature) Sensor Error Detection: Yes Sensor Error Current: 2 µA nominal (when detecting) Current Input Measurement Range: 0 to 20 mA Programmable Measurement Ranges: 0 to 20 mA and 4 to 20 mA Input Resistance: 20 Ω + PTC 50 Ω nominal 2-Wire Transmitter Supply: >15 V/20 mA Voltage Input Measurement Range: 0 to 12V Programmable Measurement Ranges: 0 to 1V, 0.2 to 1V, 0 to 5V, 1 to 5V, 0 to 10V, 2 to 10V Input Resistance: 10 MΩ nominal
OUTPUT Current Output Signal Range: 0 to 20 mA (span) Programmable Signal Ranges: 0 to 20 mA, 4 to 20 mA, 20 to 0 mA and 20 to 4 mA Load: 20 mA/600 Ω/15 Vdc max Load Stability: ≤0.01% of span/100 Ω (span=currently selected measurement range) Range Limits (NAMUR NE43 Out of Range): Below 3.8 mA or above 20.5 mA for 4 to 20 mA output; 0 mA or above 20.5 mA for 0 to 20 mA output Sensor Error Detection: Below 3.5 mA or above 23 mA for 4 to 20 mA output; 0 mA or above 23 mA for 0 to 20 mA output Current Limit: ≤28 mA Voltage Output Signal Range: 0 to10 Vdc Programmable Signal Ranges: 0 to 1V, 0.2 to 1V, 0 to 5V, 1 to 5V, 0 to 10V, 2 to 10V, 1 to 0.2V, 1 to 0V, 5 to 1V, 5 to 0V, 10 to 2V and 10 to 0V Load (Min): >10 kΩ
GENERAL Supply Voltage: 16.8 to 31.2 Vdc via power rail or connectors Power Consumption: 1.2 W max Internal Consumption: 0.4 W typical, 0.65 W max Isolation: Input/output/power Isolation Voltage (Test): 2.5 kVac Isolation Voltage (Working): 300 Vac Status LED: Green LED indicates operational status of the unit and input sensor Normal Operation: Flashes for 15 ms at 13 Hz rate Sensor Error: Flashes for 15 ms at 1 Hz rate Hardware Failure: LED off Signal/Noise Ratio: >60 dB Response Time (0 to 90%, 100 to 10%) for Temperature Input: ≤1 s Response Time (0 to 90%, 100 to 10%) for mA/V input: ≤400 ms Accuracy: Absolute accuracy or accuracy from table on PDF (whichever is greater) Absolute Accuracy: All Input Types: ≤±0.1% of span (selected input range) Temperature Coefficient: All Input Types: ≤± 0.01% of span/°C EMC Immunity Influence: <±0.5% of span Extended EMC Immunity NAMUR NE 21, A Criterion, Burst: <±1% of span (span = selected input range)
ENVIRONMENTAL Operating Temperature: -25 to 70°C (-13 to 158°F) Storage Temperature: -40 to 85°C (-40 to 185°F) Calibration Temperature: 20 to 28°C (68 to 82°F) Relative Humidity: 0 to 95% RH non-condensing Protection Degree: IP20 Installation Area: Pollution degree 2 and measurement/overvoltage category II
MECHANICAL Dimensions: 113 H x 6.1 W x 115 mm D (4.4 x 0.24 x 4.5") Weight: 70 g (0.15 lb) approx DIN Rail Type: DIN EN 60715 - 35 mm Wire Size: 0.13 x 2.5 mm²/AWG 26 to 12 stranded wire Screw Terminal Torque: 0.5 Nm
Ordering Example: (1) DRSL-ADAPTOR Configuration adaptor/docking station for use with DRSL-DISPLAY programming interface, ¥49,975 plus (1) DRSL-DISPLAY Display/programming front communication interface for DRSL-U (plugs into DRSL-ADAPTOR), ¥13,685 plus (1) DRSL-U Isolated universal input signal conditioner, ¥48,865, ¥49,975 + 13,685 + 48,865 = ¥112,525
製品マニュアル:
ダウンロード
Installation Guide Comm Interface for DRSL-U
ダウンロード
DIN Rail Signal Conditioner Universal Input