  • ±1% Accuracy (Up to 500 SLM), ±0.2% Repeatability
  • Low Cost
  • For Flowrates Up to 1000 SLM
  • Normally Closed Valve for Added Safety
  • Fast Response
  • Altitude Insensitive
  • Removable Sensing Tube Assembly

商品説明

The FMA-700A/800A series has been discontinued. Please see the FMA2700/FMA2800 and FMA3700/FMA3800 series as a possible alternative or contact our Flow Engineering department.

The OMEGA™ FMA-700A and 800A Series mass flowmeters and controllers extend the range of mass flow measurement and control from 0 to 5 SCCM to 1000 SLM with 50:1 turndown capability. Like all mass flow sensors, these new units measure mass flow rate directly, without the need for temperature or pressure corrections.

The FMA-700A and 800A Series work on the heated bypass sensor tube principle. The units feature corrosion resistant stainless steel and FKM construction and a 325 mesh (44 micron) inlet screen to prevent contamination of the sensor. The flow body and valve leak less than 10-8 atm-cc/sec of helium. In the event of a power failure, the valve is normally closed.

The FMA-78P2 and FMA-78P4 are 2- and 4-channel readout/power supply/setpoint control electronic boxes for use with FMA-700A and 800A Series mass flow sensors. The units feature 000.0 to 102.4% of full scale display, a ±15 Vdc power supply, setpoint potentiometers (for mass flow controllers), power cord, and an electrical connector for hookup of the electronics boxes’ outputs.

The FMA-78P Series rate indicators/electronics packages also have back-panel blend control switches that select independent/slave operations for each channel (channel 1 is the master), and the units include a 0 to 5 Vdc output for each channel for chart recording, data acquisition, or computer interfacing.

SPECIFICATIONS
Accuracy: ±1% full scale (up to 500 SLM), ±2% full scale for 750 to 1000 SLM
Repeatability: ±0.2% full scale
Turndown Ratio: 50:1
Response Time: 2.0 seconds
Analog Setpoint:
   0 to 5 Vdc: Voltage output models
   4 to 20 mA: Current output models
Linear Output:
   Standard: 0 to 5 Vdc into 2000 Ω min.
   Optional: 4 to 20 mA, into 200 to 750 Ω
Max Pressure:
   FMA-760A to FMA-772A: 1000 psig
   FMA-773A to 778A: 200 psig
   FMA-860A to FMA-874A: 1500 psig
   FMA-875A to FMA-878A: 1000 psig
       
Pressure Coefficient: 0.1% typical
Differential Pressure:
   FMA-800A Series: 0.5 psid
   FMA-700A Series: No less than 20% of supply pressure, depending on parameters
Temperature Range (Gas and Ambient): -10 to 70°C (14 to 158°F)
Temperature Coefficient: 0.05% full scale/°C
Wetted Parts: 316 SS body and sensing tube assembly, FKM elastomers, other materials available
POWER
Voltage Output Models: 12 Vdc (±5%) or 15 Vdc (±10%)
mA Output Models: 15 Vdc (±5%) or 24 Vdc (±15%)
Current Consumption:
   FMA-800A: <45 mA DC
   FMA-700A: <250 mA DC
注文する (部品番号ビルダーより掲載以外の製品もご選択いただけます)(注文数を入力の上、"カートに追加"をクリックしてください)

他のお客様はこのような商品も購入しています。: このアイコンをクリックすると、このモデルを購入した他のお客様の購入商品リストが表示されます.

部品番号/Desc.
注文数
商品説明
Mass Flowmeters with Compression Fitting Connections
  
FMA-864A-V
¥270,215 (税込)
(税抜価格 ¥245,650)
納期: 2 週間
 
1/4", 100 SCCM
Accessories
  
FMA-78C25
お問い合せ下さい   
納期: 
 
 
25' interconnecting cable with connectors
Click to expand/minimize a list of what others bought with theFPW-15
FPW-15
お問い合せ下さい   
納期: 
 
 
15 Vdc power supply for "A" version only
注: Comes complete with operator's manual and mating connector.


 
We noticed you are from the US. Visit your local site for regional offers and live support.