The OM-CP-PH101 is a low cost, high
resolution, miniature battery powered, stand-alone
data logger used for automatically
recording pH between 0.00 and 14.00 pH.
The OM-CP-PH101 uses a 16-bit ADC to
achieve a resolution of 0.01pH. In addition,
the OM-CP-PH101 will directly connect to
many off-the-shelf pH electrodes. The user
can easily calibrate to a specific electrode
through our software. This all-in-one compact,
portable, easy to use device will measure and
record up to 8,192 pH and 8,192 temperature
measurements.
The OM-CP-PH101 is a
major leap forward in both size and
performance. Its real time clock ensures
that all data is time and date stamped. The
storage medium is non-volatile solid state
memory, providing maximum data security
even if the battery becomes discharged. Its
small size allows it to fit almost anywhere.
Data retrieval is simple. Plug it into an available
USB port and our easy to use software does
the rest. The software converts your PC into a
real time strip chart recorder. Data can be
printed in graphical or tabular format and can
be exported to a text or Microsoft Excel file.
Specifications
TEMPERATURE CHANNEL
Temperature Sensor: 100 ΩPt RTD Calibrated Accuracy:
±0.01°C @ 25°C ambient (±0.015Ω) Temperature Resolution: 0.01°C (0.001Ω) Measuring Range:-200 to 850°C (-328 to 1562°F); 0 to 500Ω Input Connection: Removeable screw-terminal
pH CHANNEL
pH Range: 0.00 to 14.00 pH (-1000 to 1000mV) pH Resolution: 0.01 pH, (0.1mV) Calibrated Accuracy: +0.1pH (+1mV) Input Connection: Female BNC jack Input Resistance: 1012 Ω typical
GENERAL SPECIFICATIONS
ADC Resolution: 16-bits pH Calibration:Digital calibration is
available in software Recording Interval: 2 seconds to
12 hours selectable in software Real Time Recording: Device may be
used with PC to monitor and record
data in real time Green Visual Indicator: LED flashes
at selected reading rate Memory: 13,107 readings per
channel, 26,214 total readings Power: 9 V lithium battery included;
user replaceable Battery Life: 1 year typical Time Accuracy: ±1 minute per month
at 20°C Data Format:
Date and time stamped, °C, °F, K, °R, Ω; pH, V,
mV, engineering units specified
through software Computer Interface: USB (interface cable required) Software: XP SP3/Vista and 7
(32-bit and 64-bit) Operating Environment:
-5 to 50°C (23 to 122°F);
5 to 95% RH non-condensing Dimensions:
61 mm H x 115 mm W x 26 mm D (2.4 x 4.5 x 1.0") Weight: 120 g (4 oz)
注: Operator's manual is included with the OM-CP-IFC200 Windows software and USB intergface cable (software is required to operate the data logger and is sold separately). Ordering Example: (1) OM-CP-IFC200 Windows software and 3.7 m (12') USB interface cable, ¥30,882