フリーダイヤル 0120-769-103
Emailjpesales@dwyeromega.com

Fast ,Very High Temperature, Fixed Distance Infrared Sensors

OS1561, OS1571Superseded Product

Very High Temperature Infrared Sensors | OS1561, OS1571Superseded Product

Very High Temperature Infrared Sensors

スクロールダウンし、全ラインアップをご覧ください

2 year warranty
  • ±1.0% rdg Accuracy
  • ±0.5% rdg Repeatability
  • 10 msec Response, 0 to 63.2%
  • Rugged Construction
  • Adjustable Emissivity
  • Unaffected by Most Electrically Noisy Environments
  • Optional Air Purge for Dirty Environments
  • Backlight Capability Available
  • Peak Sense and Hold Standard
  • Ranges to 390°C (735°F) with Lead Sulfide Detector
  • Ranges to 3700°C (6692°F) with Lead Selenide Detector
  • 1 mV per Degree Output Standard with Optional Thermocouple, 0 to 10 Vdc or 4 to 20 mA Output Available
  • Signal Processing Available with OS1571

商品説明

The OS1561 and OS1571Series has been superseded, please see the OS4000 Series as a possible substitute or consult OMEGA's Temperature Engineering department.

The OS1561 and OS1571 stand-alone line of sight infrared pyrometers provide a 10 msec time constant for applications requiring exceptionally fast and accurate non-contact temperature measurements.

It is a compact and rugged unit, with 1 mv/degree outputs, and is compatible with a variety of temperature instrumentation including meters, recorders and dataloggers. It has an optional linear output of either 0 to 10 Vdc or 4 to 20 mA

The detector has a built-in synchronous demodulation circuitry, making it ideal for use in applications with high levels of electrical noise.

The OS1571 offers signal processing outputs of peak sense and hold. The pyrometer also has the ability for backlighting which defines the exact area being measured by projecting an image of light at a particular point on the object.

The OS1561 and OS1571 detectors are available in various temperature ranges with specific spot sizes at specific target distances (see ordering matrix on next page for details). It is intended for fixed mount applications where the pyrometer location, relative to the object being measured, will be constant.

Range Code Temperature
Range
Spectral
Response
Detector
Type
Spot Size at
Distance to Target
R1130 to 260°C (266 to 500°F)1.8-3.0 micronsLead Sulfide9.7" @120"
R2185 to 390°C (365 to 734°F)1.8-3.0 micronsLead Sulfide2.35" @ 120"
R370 to 140°C (158 to 284°F)1.8-3.0 micronsLead Sulfide0.10" @ 2.0"
R4100 to 300°C (212 to 572°F)1.8-3.0 micronsLead Sulfide0.17" @ 12.0"
R5200 to 600°C (342 to 1112°F)4.5-5.15 micronsLead Selenide9.7" @ 120"
R6800 to 2400°C (1472 to 4352°F)4.5-5.15 micronsLead Selenide2.35" @ 120"
R71000 to 3000°C (1832 to 5432°F)4.5-5.15 micronsLead Selenide0.17" @ 12.0"
R81235 to 3700°C (2255 to 6692°F)4.5-5.15 micronsLead Selenide0.49" @ 28.0"
Other temperature ranges and spot sizes vs. distance to target ranges are available; consult factory for details.

Second Output
OS1561 Models - No Add’l Cost
Optional Second Output
OS1571 Models
Order Suffix Description
-V0 to 10 Vdc
-MA4 to 20 mA
-JJ thermocouple
-KK thermocouple
Order Suffix Description
-V0 to 10 Vdc
-MA4 to 20 mA

Optional Signal Processing Feature
OS1571 Models only

Order Suffix Description
-PSPeak Sense and Hold
-SASignal Advertising
-FRFast Rise Variable Decay
注文する (部品番号ビルダーより掲載以外の製品もご選択いただけます)(注文数を入力の上、"カートに追加"をクリックしてください)
部品番号/Desc.
注文数
商品説明
Example part numbers are shown below. For additional configurations see the Part Number Builder below this table.
OS1561C-R1-J
お問い合せ下さい   
納期: 
 
 
fixed distance pyrometer with 130 to 260°C range, 1 mV/°C output, J thermocouple second output
OS1561F-R2-MA
お問い合せ下さい   
納期: 
 
 
fixed distance pyrometer with 365 to 734°F range, 1 mV/°F output, 4 to 20 mA second output
OS1571C-R1
お問い合せ下さい   
納期: 
 
 
fixed distance pyrometer with 130 to 260°C range, 1 mV/°C output
OS1571C-R1-V
お問い合せ下さい   
納期: 
 
 
fixed distance pyrometer with 130 to 260°C range, 1 mV/°C output, optional 0 to 10 Vdc second output
OS1571F-R2-MA-PS
お問い合せ下さい   
納期: 
 
 
fixed distance pyrometer with 365 to 734°F range, 1 mV/°F output, optional 4 to 20 mA second output, Peak Sense and Hold
Accessories
OS1562-MB
お問い合せ下さい   
納期: 
 
 
Mounting Bracket
PSS-D15B
お問い合せ下さい   
納期: 
 
 
±15 Vdc or 30 V regulated power supply, ±100 mA, 0.025 line and load, 115 Vac input
注: Specify range code from range chart above. Units come complete with operator’s manual.
Model OS1561 can use range codes R1 thru R4. Model OS1571 can use range codes R1 thru R8.

部品番号ビルダー

下記から仕様構成を選択し、ご注文ください
以下からオプションを選択:


(1) Model :
1561 : model 1561
1571 : model 1571

(2) Output :
C : 1 mV/°C output
F : 1 mV/°F output

(3) Range :
R1 : 130 to 260°C (266 to 500°F)
R2 : 185 to 390°C (365 to 734°F)
R3 : 70 to 140°C (158 to 284°F)
R4 : 100 to 300°C (212 to 572°F)
R5 : 200 to 600°C (342 to 1112°F)
R6 : 800 to 2400°C (1472 to 4352°F)
R7 : 1000 to 3000°C (1832 to 5432°F)
R8 : 1235 to 3700°C (2255 to 6692°F) (NOTE R5 R6 R7 R8 ranges with Lead Selenide Detector are only available on OS1571 models.)

(4) Second Output(standard with OS1561 and optional with OS1571) :
オプション選択がない場合 : no second output(OS1571 only)
-V : 0 to 10 Vdc
-MA : 4 to 20 mA
-J : J thermocouple(OS1561 only)
-K : K thermocouple(OS1561 only)

(5) Optional Signal Processing Feature(OS1571 only) :
オプション選択がない場合 : no options signal processing feature
-PS : Peak Sense and Hold
-SA : Signal Advertising
-FR : Fast Rise Variable Decay
注: 有効な部品番号については、「商品仕様(PDF)」を参照ください。
ご不明な場合は、カスタマーサービス(0120-040-572 もしくは jpesales@dwyeromega.com)へお問合せください。