SPECIFICATIONS Excitation: 12 to 36 Vdc Output: 4 to 20 mA (2 wire) Operating Temp: -29 to 85°C (-20 to 185°F) Compensated Temp: -18 to 72°C (0 to 160°F) Thermal Effects: Zero: 0.015% FS/°F Span: 0.015% FS/°F Proof Pressure: 10 psi Burst Pressure: 50 psi Static Pressure: 100 psi Gage Type: Capacitance Max Loop Resistance: (Supply voltage - 12) x 50 Ω Calibration Report: NIST calibration at 25, 50, 75 and 100% FS; upscale and downscale provided Response Time: 250 ms Wetted Parts: Dry, clean non-corrosive gases only Enclosure: NEMA 4X (IP65); 137 L x 76 W x 64 mm D (5.4 x 3 x 2.5") Pressure Port: ¼ NPTF Electrical Conn: Screw terminal (two ½ NPTF conduit) Weight: 955 g (2.1 lb)
注: Comes with complete operator’s manual. To Order the High Accuracy PX657, change PX656 to PX657 in Model Number. See Part Number Builder below. Ordering Example: (1) PX656-01DI is a high accuracy, 0-1 inH2O differential transmitter with NEMA-4 enclosure and 4-20 mA output, ¥194,740
部品番号ビルダー
下記から仕様構成を選択し、ご注文ください
以下からオプションを選択:
(1)
Accuracy : 6 : Standard Accuracy 7 : High Accuracy